This assay is intended as an aid in the diagnosis of inflammatory bowel disease (IBD), specifically Crohn's disease. The kit is designed for the detection and quantitation of IgA anti-Saccharomyces cerevisiae antibodies in human serum.
Antigen
The microwells are precoated with mannan isolated from Saccharomyces cerevisiae.
Calibration
The assay is calibrated to U/mL against an arbitrary internal reference calibrator, as no internationally recognised reference preparation is currently available. A 5 point calibration curve is generated. A cut-off control is provided if being used as a screening assay.
Result Interpretation
Quantitative Assay
| Result | Interpretation |
|---|---|
| < 7.0 U/mL | Negative |
| > 7.0 U/mL | Positive |
Screening Assay
| Result | Interpretation | Action |
|---|---|---|
|
OD greater than cut-off control
|
Suspected of having ASCA
|
Test in the quantitative assay if value required |
| OD less than or equal to cut-off control | Negative | Report as negative |
Measuring Range
1.23 - 100 U/mL
Sensitivity
0.93 U/mL